Equipment log for the Zeiss SEM Ultra60
| 2009-7-2 | Image noisy.
took another look at the Ultra60 and the images seem better. Maybe it just took some time for the tip to stabilize.
The images are still pretty soft and they seem noisy so we would still like you to come and take a look at our system.
I took a few measurements.
Specimen current (Faraday Cup, 5mm WD, 30 痠 aperture):
5KV = -158pA
10KV = -201pA
Emission current:
EHT off = 318 渙
EHT on = 256 渙 |
| 2009-6-9 | The service engineer was here today working on the tool. |
| 2009-10-8 | The system is down. A user hit the pole piece with the sample stage. |
| 2009-10-7 | The system is down. Further information will be supplied as it becomes available. |
| 2009-10-14 | Stigmation out of range
|