Equipment log for the Karl Suss MA-6 Mask Aligner


2009-7-9 Back-Side Alignment
The issue with the power supply has been resolved.
2009-7-7 Back-Side Alignment
Power supply went bad. Problem will be further investigated in the morning.
2009-7-22 Back-Side Alignment
The MA6 is back up.
2009-7-21 Back-Side Alignment
The problem that occurred today with the machine not turning on is the product of either a bad relay or a power supply. We should have that part in stock. The tool will be examined tomorrow and the replacement part should be installed.
2009-7-20 Back-Side Alignment
The problem that occurred today with the machine not turning on is the product of either a bad relay or a power supply. We should have that part in stock. The tool will be examined tomorrow and the replacement part should be installed.
2009-7-17 Back-Side Alignment
The problem that occurred today with the machine not turning on is the product of either a bad relay or a power supply. We should have that part in stock. The tool will be examined tomorrow and the replacement part should be installed. Thank you.
2009-7-16 Back-Side Alignment
Power supply / relay issue will be addressed by staff tomorrow.
2009-7-14 Back-Side Alignment
The problem that occurred today with the machine not turning on is the product of either a bad relay or a power supply. We should have that part in stock. The tool will be examined tomorrow and the replacement part should be installed. Thank you.
2009-6-24 Back-Side Alignment
If you are having difficulties with and/or inconsistent results with your lithography processes, please contact the following processing team members: Janet Cobb-Sullivan or Mikkel Thomas.
2009-6-23 Information
If you are having difficulties with and/or inconsistent results with your lithography processes, please contact the following processing team members: Janet Cobb-Sullivan or Mikkel Thomas.
2009-6-18 Back-Side Alignment
The issue with the overcurrent issue caused by a misposition backside alignment camera has been fixed by a manual realignment which then reinitialized the system. The system is working properly.
2009-6-11 Miscellaneous
If you are instructed by staff to turn off the machine and turn it back on, DO ***NOT*** turn off the power supply underneath the tool. Additionally, if you are instructed by staff to restart the system, make *SURE* that you know what should appear on the screen for it to confirm that it will be working properly.
2009-5-27 Lamp Power Supply
A user reported the tool alarming because the power supply was providing too much power to the lamp. The power supply has been adjusted and tool has been recalibrated. If the power supply starts alarming, DO NOT CONTINUE USING THE TOOL. Immediately stop, put a sign on the tool saying machine down, and call the staff. Continued use of the machine in that situation is a violation of cleanroom policy.
2009-5-26 Back-Side Alignment
The Karl Suss MA-6 Mask Aligner has been marked down by Eric Woods for the following reason: The lamp power supply was alarming for CI2. The power level was too high. The power supply was adjusted to bring the power back into compliance, but it would not stabilize and drifted one way or another. The tool will be further examined tomorrow.
2009-5-19 Calibration
The system has been calibrated. The new intensities are 5mW/cm2 for CI1 and 15mW/cm2 for CI2. All the optical components have been cleaned. The system is now once again operational. If you have any questions or concerns regarding this matter, please contact the staff.
2009-5-12 Back-Side Alignment
System does not support proximity mode. it is not available.
2009-4-23 Back-Side Alignment
The tool is operational. If you experience any problem with the tool, please contact the staff.
2009-4-21 Information
The tool is down because the EM general power safety relay still needs to be replaced. Additionally, there is an issue with the microscope position sensor. These issues are being investigated and will hopefully be resolved soon. The changeover of the TSA to DUV will not occur until the BSA MA6 is back online and will be scheduled at that time.
2009-4-20 Miscellaneous
A safety relay for the emergency shutoff sub-system has failed. This relay controlled power to the entire tool. This relay is currently being sourced and a replacement will be ordered tomorrow.
2009-4-15 Back-Side Alignment
Users are NOT allowed to cut blue tape on the mask holder as this has resulted in the seal on the vacuum chuck getting cut. This is expensive to replace and it takes the machine down for a significant length of time.
2007-7-3 Miscellaneous
We are working on the loss of wafer vacuum error and will send out an email when it is fixed.
2007-6-21
The problem with the wafer vacuum is still being worked on. We hope to resolve the issue soon.
2007-5-3
We are continuing to work on fixing the backside alignment. If you really need to do backside alignment, please contact Vinnie to discuss other available options.
2007-5-1
The left-hand side backside alignment microscope is down. Repairs are in progress.
2007-4-2
A user reported an overcurrent error over the weekend. This morning we found that the overcurrent error was no longer there. Users are not permitted to try and fix any problems with machines in the cleanroom. The overcurrent error was due to a user error when using the back side alignment. Please be aware of the position for the back side microscope.
2007-3-13
The MA-6 is back up. If you have any problems, please contact Vinnie.
2007-3-12 Miscellaneous
The Z-axis motor sensor is in the process of being repaired. We will email the MA6 users when the machine becomes available again.
2007-2-23
The topside alignment microscope has been repaired. The loose bearing on the base has also been fixed.
2007-2-2 Information
The 360nm filter for the mask aligner is now available. The filter must be checked out from the staff. Please contact Vinnie if you need to use this filter.
2007-10-23
If you are using the Mask Aligner, and notice a "Z-axis over current error", please do not try and restart the system. Instead, please call Vinny immediately.
2007-10-12
The MA-6 is back up. Thank you for your patience while we fixed it. If you notice any problems, please let a staff member know immediately.
2007-10-11
Please do not adjust the intensity of channel 2 on the MA-6. If you have any questions, please contact Vinnie.
2007-1-29 Information
A user reported overcurrent error. If this problem occurs, please report this problem to a staff member immediately. Do not turn the machine on and off, and do not try to fix it yourself.
2007-1-25 Lamp Replacement
The lamp on the MA6 has been changed and the intensity for Cl1 is 5 mW/cm2 and 22 mW/cm2 respectedly. Please adjust your exposure time accordingly.
2006-9-8
A user complained that the mask holder shifted during exposure. We replaced the mask holder frame, which should fix the problem. Please inform the staff if the problem still occurs.
2006-9-27
The tool has been reset and should be fully operational. If you have any problems, inform staff immediately.
2006-9-26
A user reported problem with the topside alignment microscope. The system was reset this morning. If the users discover that the microscope is not moving in the x or y direction, please check to see if the BSA button has been pressed. Note that the button was accidentally pressed and this was the cause of the problem.
2006-9-22
There is a problem with the stage; if you see a z-axis, call Vinny and he will tell you how to fix the problem. We correct the problem in the near future.
2006-6-12 Calibration
This is a reminder to users. Please be sure to use the proper probe when measuring the intensity of the lamp.
2006-5-31 Information
A user reported a over current error on the system. The error was cleared, and the system was made available again.
2006-5-15 Information
The system was logged down for a short period of time on Friday due to a problem with the back side aligment mode. If any users want to do back side alignment, please contact Vinnie.
2006-3-17 Information
A user reported a problem with the mask vacuum. The access controller was not working properly. We will replace the relay on Monday. In the mean time, if you are logged in, just tap the box to get the vacuum sensor to work. If this does not work, contact the staff.
2006-1-26 Information
The 360nm filter was delivered today. It is now available to all users. Please remember to be careful when handling the filter.
2005-8-8 Information
A user reported a problem with one of the alignment controllers. They reported that X direction would not work. We found that the controller was working properly. This may be an intermittent problem. If any users experience this type of problem, please report it right away.
2005-6-3
A cleanroom user reported a problem with the X alignment of the MA-6 stage. This problem will be resolved sometime next week.
2005-6-23
A cleanroom reported that there was a problem with movement of the stage. The x-axis of the stage have been readjusted and once again works as intended. Remember, the stage does have limits to how far it can move in both the x and the y direction. If you exceed these limits you can cause damage to the stage.
2005-5-2
Over the weekend a cleanroom user reported that their sample was shifting during alignment. In order to remedy this adjustments can be made to the WEC pressure setting on the control panel of the MA-6. Also, if this problem arises during your alignment check to make sure that the WEC head is clear of any obstructions that could cause a vacuum loss. In this case a piece of blue tape was found on the edge of the WEC head. If you have any questions about obstructions to the WEC head or WEC pressure setting please contact Vinny.
2005-5-18
The mask holder and chucks have been relocated for the MA-6 Mask Aligner. Please be sure to properly store all chucks in their newly designated locations.
2005-5-16
We fixed the MA6 problem today. There were several issues about the machine that needed to be addressed to all the users. First, we DO NOT ALLOW users to TURN OFF and TURN ON the machine without MiRC staff's permission. If you do so, your access to the machine will be cut off, no exception. Second, the PROXIMITY mode on the machine is not available right now. If you want to use it, you need to contact Vinny and do it under his supervision. Finally, all users MUST KNOW how to change the contact mode such as Soft contact, Hard contact, Vacuum and Low vacuum contact to suit with your exposure process. If you are not quite sure how to do it, do not hesitate to request for a full training or check off on the tool again.
2005-4-8
A cleanroom user reported a vacuum problem occuring while they were running the system. Due to the intermittent state of the vacuum we will be scheduling some down time next week to take a more in depth look at the situation.
2005-4-7
Today a cleanroom user reported a poor vacuum error on the MA6 mask aligner. Upon investigation we found that photoresist had been left on the vacuum chuck and the WEC head causing a leak in the seal. The photoresist was removed and system was restored.
2005-4-26
There seems to be some minor confusion regarding how to properly change out the new mask holder setup. If you have any questions or concerns regarding this subject please contact Vinny.
2005-4-20
When using the new mask holder be sure not to tangle the vacuum line while you are attempting to operate it.
2005-4-15
A new vacuum line and connection was installed yesterday. If you are unsure of how to properly change out mask holders with the new connection in place please contact Vinny for directions.
2005-4-14
Today the MA-6 mask aligner had its vacuum lines and connection replaced, resolving all the vacuum issues that have been occuring on this system. If you notice that you mask is still being dropped please contact MiRC immediately.
2005-4-13
The MA6 Mask Aligner will be log down from 10am tomorrow until further notice. We will replace the vacuum fitting on the machine in order to work with vacuum fitting on the new mask holder. Please schedule to use the machine accordingly.
2005-4-12
The MA6 will be down tomorrow as staff performs some maintenance and premaintenance to ensure the proper functionality of this system.
2005-3-9
The lamp for the MA-6 Mask Aligner will be replaced first thing tomorrow morning. Users are reminded that should periodically check the intensity of the lamp with the intensity probe, any intensity fluctuations should be reported to MiRC Staff.
2005-3-28
Today, a cleanroom user reported that there was a weak vacuum present at the mask holder. A kink and small hole was detected in the physical vacuum line causing this weak vacuum. The line was replaced, solving this problem. If you notice any additional problems with vacuum or this system in general please contact MiRC Staff.
2005-3-22
If any users wish to use the backside alignment feature please contact Vinny in order to get proper instructions.
2005-3-2
The MA6 has been fixed and back was restored to normal operation. If you have troubles to align your wafer, please contact Vinny.
2005-3-10
The lamp on the MA-6 mask aligner has been replaced today. Everyone should be testing the intensity of the new lamp in order to make appropiate changes to their exposure times.
2005-3-1
If any MA-6 users wish to use proximity mode or low vac/vac mode please contact Vinny.
2005-2-8
The MA-6 Mas Aligner was down today as MiRC staff fixed a relay problem that was occuring. Because of this the focus for the top substrate is once again working. Users are reminded that if you are having a problem with vacuum/low vacuum mode to please contact Vinny.
2005-2-7
MiRC staff will be fixing the relay problem that has been occuring on the MA-6 mask aligner, this will cause the machine to be marked down from 8 until 12 tomorrow morning.
2005-2-4
We have discovered the source of the problem. We are in the process of ordering a replacement part. For now, the right focus for the top substrate on both top and backside alignment is still in intermittent mode.
2005-2-3
We have installed the potentiometer for the right focus knob. There is still a problem with the focus. Will continue investigating the problem tomorrow.
2005-2-28
Today, a cleanroom user reported a problem with the 4" mask holder's vacuum. The vacuum problem was resolved and the system is once again ready for use. Remember when you are loading your mask that it has been properly cleaned as this could cause a lack of vacuum on your mask. If any user has any problems please contact MiRC Staff.
2005-2-2 Information
The potentiometer came in late today. We will install it onto the machine tomorrow.
2005-2-11
We have made some adjustments to the left backside microscope. Users should no longer have any more problems. If you discover any more problems, please contact Vinnie.
2005-2-1 Back-Side Alignment
The focus on the back-side mircoscope has been fixed. If any additional problems occur in the system please contact Vinny.
2005-11-9 Information
A user reported that the 365nm filter had been damaged. We are currently in the process of ordering a new filter. We will notify all MA6 users when the new filter arrives.
2005-1-7 Maskholder
A cleanroom user reported that the 3 inch chuck has produced errors during the WEC performance step. Upon investigation we found that one of the mask holder clips was actually thicker then the others. This odd clip was replaced with a proper sized one and the errors that were occuring are now resolved.
2005-1-5 Information
Please do not adjust the WEC pressure if you do not know how to do it properly. We have discovered that some users have been adjusting the pressure too low.
2005-1-4 Information
A cleanroom user reported the chuck not properly contacting his wafer. We will be looking into this problem tomorrow.
2005-1-31 Back-Side Alignment
The MA-6 mask aligner will be brought down from 10-12 tomorrow morning while its back-side microscope is being repaired.
2005-1-27
Today, a cleanroom user reported that when they attempted to grab an image the left portion of the image was not as sharp as the right. MiRC is currently investigating this issue. Look for a future email for further updates.
2005-1-26
A cleanroom user reported troubles while attempting to focus on their sample. Upon further investigation we found that the light bulb for the microscopes had burned out. The bulb was change and the system is once again focusing properly.
2005-1-14 Maskholder
If any user wants to use the vacuum/low vacuum setting for the 5 inch mask holder you will have to contact Vinny first.
2005-1-11
Today, a cleanroom user reported that they were having problems with the 5 inch mask holder not being able to properly hold their mask. We later found that the thickness of this user's mask was twice the size of the normal masks. Please be mindful of this problem when you are having your masks made, and provide the appropriate dimensions.
2004-12-8 Maskholder
We are in the process of ordering a new 5" maskholder. However, it will take at least a month for the new maskholder to come in. Until then, please remember that the old maskholder has been damaged, and thus will not hold the masks in place all of the time.
2004-12-7 Maskholder
The 5" mask holder has been damaged, and is not level anymore. If you attempt to use either the Vacuum or the Low Vacuum modes, then there is a 50% chance that the maskholder will not hold the mask. We will continue to look into this issue tomorrow.
2004-12-3 Lamp Power Supply
A cleanroom user reported that the lamp power was off. We have restarted the lamp power supply, and the system is functioning properly. Please report any further issues to a cleanroom staff member.
2004-12-14
We have placed the order for the new 5" mask holder; however, it may take up to several months to arrive.
2004-12-10
Several cleanroom users have reported issues with the 5" mask holder. We are in the process of ordering a new mask holder; however, please be aware that it will take 8 to 12 weeks for the new mask holder to arrive. We apologize for any inconvenience.
2004-11-8 Maskholder
We have installed new clamps onto the 4" and 5" mask holders. From now on, the clamp will stay in the compressed position once you pull it back. After you turn on the vacuum, press the metal piece to release the clip.
2004-11-30 Maskholder
Due to the problems with the ICP, we were unable to work on the mask holder vacuum issue today. However, we will fix this problem in the coming days.
2004-11-3 Maskholder
Tomorrow, we will replace the faulty clamp on the 5" mask holder. This will improve the performance of the mask holder.
2004-11-29 Maskholder
The MA-6 will be down tomorrow. We will be working on the mask holder vacuum problem. We will send out an e-mail when the MA-6 is logged back up.
2004-11-22 Maskholder
A cleanroom user reported issues with the mask vacuum. We will look into this problem tomorrow.
2004-11-1 Maskholder
If you lose vacuum while trying to load a mask onto a mask holder, try mounting another mask before contacting the MiRC staff. This will determine if the problem is with the mask holder or the mask. If you still receive vacuum errors, then notify a staff member.
2004-10-8 Back-Side Alignment
Backside alignment is currently unavailable on this machine. We have removed the right backside microscope for further testing. We will send out an e-mail once we have completed these tests and re-installed the microscope onto the system.
2004-10-28 Information
Cleanroom users are reminded that NO user is allowed to restart the MA-6 mask aligner. Any questions about this policy should be directed to Gary Spinner.
2004-10-18 Information
We have been experiencing an intermittent problem with the top substrate right focus knob. Today, we spoke to Karl Suss about this issue. The MA-6 will be down tomorrow for maintenance tomorrow from 11:00 AM until 1:00 PM as we address this issue.
2004-10-11 Back-Side Alignment
We examined the system today, and the backside alignment microscopes are now working properly. However, if you have any difficulties performing a backside alignment, please contact Vinnie.
2000-3-3 Miscellaneous
A user reported an "OVERCURRENT" error message. We will be reviewing this problem further.
2000-3-10 Miscellaneous
Several people have experienced "OVERCURRENT" errors while loading wafers into the aligner. We have determined that this occurs when using a "Proximity" program or a program configured for "Proximity WEC" if the wafer is not centered in the machine. After the wafer is loaded, the aligner will press the wafer against the mask to make them parallel. Then it will lower it slightly for alignment. If you are using a "Proximity" program or certain other programs configured for "Proximity WEC" the aligner will extend the metal flags on the maskholder between the wafer and the mask before this process. The flags have metal balls in them that should separate the wafer from the mask. If the wafer is not centered, the balls will not be between the wafer and mask and errors will occur. Please see MiRC staff for help if you experience error messages or have trouble moving the sample during alignment.
2000-3-1 Miscellaneous
A user reported several error messages while operating the MA-6. This morning the system was working well. If you experience any more problems, please report it to MiRC staff.
2000-2-22 Maskholder
We have replaced the missing screw that secures that maskholder tray to the aligner.
2000-2-1 Back-Side Alignment
The MA-6 mask aligner would not start properly because the back-side alignment proximity flags were out of alignment. As a result of the mis-aligned flags, the back-side alignment mechanisms were running into each other causing an overcurrent error. The maskholder tray has been replaced. The new tray is made of stainless steel instead of plastic, so it should be stronger.
2000-1-31
The mask aligner is reporting an overcurrent error upon startup. We will investigate this more tomorrow
2000-1-26
The damaged mask holder tray will be replaced with a new aluminum tray. The aluminum tray will be installed as soon possible.
2000-1-21 Miscellaneous
A user reported a problem with the wafer holder slide. The holder tray had come off of its track and the track had become jammed under the maskholder. This problem has been repaired.
2000-1-14 Miscellaneous
The maskholder table was broken off a few days ago. It has been fixed.
1999-9-2 Miscellaneous
A cleanroom user reported that the system was experiencing an overcurrent condition when the chuck was placed into contact mode. The problem was traced to broken and bent limit flags that are attached to the mask holder. New flags have been ordered but ALL users will require recertification by James Zhou before they may use the aligner.
1999-8-5 Miscellaneous
A user reported that the MA-6 would not expose. A message was being diplayed that the Mirror House was not moving properly. We located a cable that was preventing the Mirror House from moving. The mask aligner is now moving properly.
1999-8-25 Miscellaneous
Several cleanroom users are experiencing problems with back-side alignment. We have spoken to a Karl Suss equipment engineer. He will be shipping us a replacement EEPROM chip on Monday of next week. We are aware of the focusing problem on the left microscope.
1999-8-13 Back-Side Alignment
The mask aligner back side-alignment scopes will not move close enough together to perform back-side alignment on the smallest wafer chuck. We are continuing to work with Karl Suss to find a solution to this problem.
1999-7-29 Back-Side Alignment
1) A user reported a "OVERCURRENT" error while using the back side alignment lamp housing. We have spent several hours examining the mask aligner and have spoken with Karl Suss about the problem. We do not have time to fix the problem at this time. Users should not move the right back side alignment lamp housing more than 50 mm left. The current position is shown on the mask aligner front panel display when the back side alignment lamp is being moved. 2) A user reported a "OVERCURRENT" error after loading his sample. We have contacted Karl Suss about the problem. There is a problem with the maskholder that is preventing the mask aligner from performing proximity alignment. We are ordering replacement parts for the maskholder. Until the 4" maskholder has been fixed it can not be used for proximity alignment.
1999-7-27 Back-Side Alignment
A user reported that the mask aligner displayed a motor overcurrent error when the right back side alignment camera was moved too far to the left. We suspect that a limit switch is not set up correctly and the alignment camera is being allowed to move further than it should. We hope to have the mask aligner back up tomorrow.
1999-7-23 Miscellaneous
A cleanroom user reported misconfiguration of software on computer. We discovered that the software was misconfigured by a user. Please do not change the configuration with out MiRC Staff approval. We will be investigating the overcurrent alarm on Monday.
1999-5-26 Maskholder
A user reported that the mask vacuum was not working. We found that one of the vacuum lines was kinked. We corrected the problem and the mask aligner is functioning properly.
1999-3-9 Information
There is now a Intensity log notebook located on the equipment. Please fill in the information when measuring the intensity.
1999-3-5 Lamp Replacement
The UV lamp for the Mask Aligner has not been replaced since the system was installed. According to the manufacturers of the equipment, the lamp must be replaced every 1000 hours in order to prevent optic damage. We, therefore, replaced the lamp and ordered a replacement. The intensity has been measured to be 21.6.
1999-3-2 Miscellaneous
The MA-6 was making a hissing sound when it was loaded. We corrected the problem. The vacuum knob on the front panel was not turned far enough. If you have a problem with a hissing sound, try adjusting the vacuum knob.
1999-3-16 Calibration
We checked the light intensity on the mask aligner.
1999-2-8 Back-Side Alignment
Problem: The leftside lamp was not burning. Solution: We replaced the backside alignment lamp and adjusted the intensity on the front of the equipment.
1999-2-5 Back-Side Alignment
The lower right lamp to the equipment was not working. We contacted Karl Suss and are continuing to work on this problem.
1999-2-3 Miscellaneous
We recalibrated the machine.
1999-2-26 Miscellaneous
The MA-6 was displaying an error message when it initialized after being turned on. We corrected the problem with the MA-6 not initializing correctly. We will be resolving the problem with the hissing nitrogen on Monday.
1999-10-6 Maskholder
New proximity flags for the four inch maskholder have arrived. We have installed them and performed several tests. If you experience any problems with the proximity flags or overcurrent errors, contact MiRC technical staff immediately. New flags cost approxatately $500 and take four to six weeks to get.
1999-10-4 Miscellaneous
The proximity flags have been received and installed. This has corrected the overcurrent problems. We are now requiring users to check out the mask holder from Gary Spinner and return it promptly.