Equipment log for the Tencor Alpha-Step Profilometer


2009-7-2 Calibration
The hard drive issue has been fixed and the tool is operational.
2009-4-20 Software
The monitor on the system failed and has been replaced. The tool is not booting properly. The problem is under investigation and an attempt will be made soon to solve it.
2007-8-16
The video cable has been swapped out and now works. The intensity is still a bit low, but the bulb will be changed out tomorrow.
2006-9-8
A user reported that the tool was not working. We honed the z-axis, tested the tool, and it is now working.
2006-6-13 Information
We have finally resolved the issue with the scanning problem we have had over the last week. The current problem we are having now is related to the stylus force. We are unable get consistent stylus force measurements. We will continue to work on this tomorrow.
2006-6-12 Information
We have had several problems with the profilometer. The latest problem involves the scanning mechanism. We have the spare parts, but we are having trouble aligning the system to scan properly. We will continue to work on it tomorrow.
2006-5-31 Information
The profilometer has been logged down due to an internal problem. One of the ribbon cables has been severed. We will try to fix the problem tomorrow and log it back up as soon as possible.
2006-5-30 Information
A user reported a problem with the homing sensor on the system. A mechanism inside of the stepper needed to be repaired.
2006-4-11 Information
The profilometer has been moved to the class 10 Photolithography bay. It can be found on the right side of the bay as soon as you enter.
2006-3-2 Information
Two problems with the profilometer have been reported. There is a problem with the light source in the system. There is a loose connection inside the system. We may need to log the system down to find the loose wire. The second problem has to do with the measurement going "out of range." We ran a stylus force calibration and reset the system. This should resolve the problem. If any users run into this problem after today, please let us know.
2006-3-10 Information
We recommend that all current users get training on the new P-15 Profilometer located in the class 1000 PL bay. The new system does a better job of measuring your samples and has a few more option available.
2005-9-29 Software
A user reported that the computer had rebooted. Please contract staff if there are any problems with the equipment. Please do not try to fix the problem yourself.
2005-7-20
The stylus on the Alpha-Step profilometer had its force adjusted to ensure that the stylus remains in contact with the sample. If you notice that the stylus is not remaining in contact please notify Chris Holshouser.
2005-6-22
Reminder to all alpha-step users: you are not allowed to adjust either of the knobs located inside the plastic shielding. If you are caught making any self adjustments your access to this equipment will be removed.
2005-6-1
The Tencor Alpha-Stepper has been reinitialized and had its stylus force adjusted. These corrections helped fixed the measuring problems that were occuring.
2005-5-30
It seems that someone either reconfigured or reinitialized the stage incorrectly on the Alpha-Step Profilometer. Staff is still in the process of realigning and reconfiguring the stylus so that the system can work as intended.
2005-5-10
Staff fixed a problem with the system's hard drive today. A system pci card had not been connected properly, but the problem has been resolved and the system is now fully operational.
2005-1-5
A user reported a problem with the camera on the profilometer. There was a loose connection to the camera inside the system. If the image does not appear again, please contact staff.
2005-1-10
The problem involving the intermitten state of the camera in the Tencor Alpha Stepper has been resolved. We found a loose wire connecting the camera to the computer; the rogue wire has been dealt with. If you still continue to experience problems with the Alpha Stepper please contact MiRC Staff.
2000-3-24 Stage
There seems to be a problem with the computer loosing track of the stage height and refusing to lift the stage high enough. We will be looking into this problem next week.
2000-1-5 Miscellaneous
A cleanroom user noticed that the diamond stylus tip was missing from the stylus and the Alpha-Step was not displaying good profiles. The stylus has been replaced and the profilometer is working well now.
1999-7-6 Stage
The stage on the Alpha-Stepper has been repaired. Do not force the stage up too far. Forcing the stage up too far causes the stylus to become misaligned.
1999-7-27 Miscellaneous
A user reported that the stylus force was too high. We lowered it.
1999-7-2 Stage
The platform on the Alpha Stepper needs to be repaired. The stepper will not go up or down consistently. It will be looked at on Tuesday.
1999-7-12 Software
A cleanroom user reported that the system kept requesting a file name after each measurement. The system software configuration has been changed to take care of this annoying problem.
1999-4-30 Miscellaneous
A cleanroom user reported that the stylus tip has material buildup on the end. We cleaned the stylus.
1999-3-9 Information
We measured film thicknesses of various samples.
1999-3-12 Information
We measured the thickness of various test samples.
1999-2-8 Information
We used the equipment to analyze various samples.
1999-2-5 Stage
We measured the thicknesses and indices of refraction of samples ran in the sputterers and those ran in the PECVD. Though the system is presently funtioning properly, we spoke with Mathew Neil, a technician from Tencor, concerning the reoccurring situation with the stage movement. He made a suggestion for a permanent solution.
1999-2-4 Calibration
We measured the standard, and the results demonstrated that the system is functioning properly.
1999-2-25 Calibration
We calibrated the equipment and measured some test samples.
1999-2-24 Calibration
We calibrated the equipment and measured various samples.
1999-2-23 Information
We calibrated and measured various samples. Also, a user was trained on the equipment.
1999-2-19 Information
We used the equipment to measure the samples ran in the E-Beam and obtained satisfactory results.
1999-2-18 Calibration
We measured various samples and calibrated the equipment. We also trained users on how to correctly measure samples with this machine.
1999-2-15 Calibration
We calibrated the equipment and measured various samples.
1999-2-12 Calibration
We calibrated and measured the equipment.
1999-2-11 Calibration
We performed calibration test on the Alpha-Step.
1999-2-1 Stage
The y-axis movement was functioning improperly. The problem has been corrected.
1999-1-7 Stage
The Y-axis control had a new problem. It has been fixed. The Alpha Step is now operable.